Most spec sheets I receive from other factories are 30 lines long and answer two of my six questions. Here is the order I read them in.
One: needle length and protrusion, as separate numbers. Length is the moulded needle; protrusion is how much stands proud of the base after assembly. A sheet that gives only one number is hiding the harder one.
Two: tolerance, not just the nominal. "0.30 mm" is a wish. "0.30 mm ± 0.02" is a commitment. No tolerance means no way to reject a batch, and you will be arguing about feel.
Three: needle count and pattern. 20 versus 40 changes the depth of penetration at the same force. If the sheet lists a count but no pattern drawing, ask for the drawing.
Four: material and plating thickness. Steel grade plus micron thickness of plating. "Surgical steel, gold plated" is a description, not a spec.
Five: sterilization method and release basis. EO with dosimetric release, or irradiated? Different shelf life, different paperwork, different logistics.
Six: pull-out force and drop test. These two are what your clinic will actually complain about. Anyone who has tested them will be proud to show the number.
Read those six and you have read the sheet. The rest is decoration, and decoration is cheap.